====== Record Data Before a Trigger Event ====== | ====== Record Data Before a Trigger Event ====== | ||
+ | The Spider has the ability to perform Circular Recording. Using this feature, you are able to save and record measurements from before a trigger event occurs.\\ | ||
+ | |||
+ | To set this up, go to **Setup -> Test Configurations -> Save/ | ||
+ | {{: | ||
+ | |||
+ | Select to begin recording "when input signal is triggered according to trigger condition" | ||
+ | {{: | ||
+ | |||
+ | Under **Recording Options**, the circular recording parameters can be set up. Check the " | ||
+ | |||
+ | //Record Duration//- The total (pre-trigger + post-trigger) recording time.\\ | ||
+ | |||
+ | //Recording Size//- Size of the pre-trigger recording.\\ | ||
+ | |||
+ | //Recording Length//- The duration of the pre-trigger recording.\\ | ||
+ | {{: | ||
+ | |||
+ | In the example seen above, when the trigger is met the Spider will record 20 seconds of pre-trigger data and then 40 seconds of post-trigger data for a total of 60 seconds.\\ | ||